欢迎访问ic37.com |
会员登录 免费注册
发布采购

5962-8872101PA 参数 Datasheet PDF下载

5962-8872101PA图片预览
型号: 5962-8872101PA
PDF下载: 下载PDF文件 查看货源
内容描述: DUAL非常低的噪声精密运算放大器 [Dual Very Low Noise Precision Operational Amplifier]
分类和应用: 运算放大器
文件页数/大小: 16 页 / 516 K
品牌: ADI [ ADI ]
 浏览型号5962-8872101PA的Datasheet PDF文件第6页浏览型号5962-8872101PA的Datasheet PDF文件第7页浏览型号5962-8872101PA的Datasheet PDF文件第8页浏览型号5962-8872101PA的Datasheet PDF文件第9页浏览型号5962-8872101PA的Datasheet PDF文件第11页浏览型号5962-8872101PA的Datasheet PDF文件第12页浏览型号5962-8872101PA的Datasheet PDF文件第13页浏览型号5962-8872101PA的Datasheet PDF文件第14页  
OP270  
NOISE MEASUREMENTS  
Noise Measurement — Noise Voltage Density  
Peak-to-Peak Voltage Noise  
The circuit of Figure 8 shows a quick and reliable method of  
measuring the noise voltage density of dual op amps. The first  
amplifier is in unity-gain, with the final amplifier in a noninverting  
gain of 101. As noise voltages of each amplifier are uncorrelated,  
they add in rms fashion to yield:  
The circuit of Figure 6 is a test setup for measuring peak-to-peak  
voltage noise. To measure the 200 nV peak-to-peak noise specifica-  
tion of the OP270 in the 0.1 Hz to 10 Hz range, the following  
precautions must be observed:  
1. The device has to be warmed up for at least five minutes. As  
shown in the warm-up drift curve, the offset voltage typically  
changes 2 mV due to increasing chip temperature after power-up.  
In the 10-second measurement interval, these temperature  
induced effects can exceed tens of nanovolts.  
2
2
Ê
ˆ
eOUT = 101  
e
+ e  
(
)
(
)
Á
Ë
˜
¯
nA  
nB  
The OP270 is a monolithic device with two identical amplifi-  
ers. The noise voltage density of each individual amplifier will  
match, giving:  
2. For similar reasons, the device has to be well shielded from  
air currents. Shielding also minimizes thermocouple effects.  
2
Ê
ˆ
eOUT = 101 2en = 101 2en  
Á
˜
3. Sudden motion in the vicinity of the device can also “feed  
through” to increase the observed noise.  
(
)
Ë
¯
4. The test time to measure noise of 0.1 Hz to 10 Hz should not  
exceed 10 seconds. As shown in the noise-tester frequency  
response curve of Figure 7, the 0.1 Hz corner is defined by  
only one pole. The test time of 10 seconds acts as an additional  
pole to eliminate noise contribution from the frequency band  
below 0.1 Hz.  
R1  
R2  
100  
10k  
1/2  
OP270  
+
eOUT  
1/2  
TO SPECTRUM ANALYZER  
en  
OP270  
+
eOUT (nV/ Hz) =101 (  
2
)
100  
80  
60  
40  
20  
0
V
= 15V  
S
Figure 8. Noise Voltage Density Test Circuit  
R3  
1.24k  
R1  
5⍀  
R2  
100k⍀  
OP270  
DUT  
enOUT  
OP27E  
+
TO SPECTRUM ANALYZER  
+
R5  
8.06k⍀  
R4  
GAIN = 10,000  
= ؎15V  
200⍀  
V
S
0.01  
0.1  
1.0  
10  
100  
FREQUENCY (Hz)  
Figure 9. Current Noise Density Test Circuit  
Figure 7. 0.1 Hz to 10 Hz Peak-to-Peak Voltage  
Noise Test Circuit Frequency Response  
Noise Measurement — Current Noise Density  
The test circuit shown in Figure 9 can be used to measure cur-  
rent noise density. The formula relating the voltage output to  
current noise density is:  
5. A noise-voltage-density test is recommended when measuring  
noise on a large number of units. A 10 Hz noise-voltage-density  
measurement will correlate well with a 0.1 Hz to 10 Hz  
peak-to-peak noise reading, since both results are determined by  
the white noise and the location of the 1/f corner frequency.  
2
Ê
ˆ2  
enOUT  
G
- 40 nV/ Hz  
Á
Ë
˜
¯
(
)
in =  
6. Power should be supplied to the test circuit by well bypassed  
low noise supplies, e.g., batteries. They will minimize output  
noise introduced via the amplifier supply pins.  
RS  
where:  
G = gain of 10,000  
RS = 100 kW source resistance  
–10–  
REV. C  
 复制成功!