欢迎访问ic37.com |
会员登录 免费注册
发布采购

3B37-B-06 参数 Datasheet PDF下载

3B37-B-06图片预览
型号: 3B37-B-06
PDF下载: 下载PDF文件 查看货源
内容描述: 隔离,热电偶输入 [Isolated, Thermocouple Input]
分类和应用:
文件页数/大小: 8 页 / 336 K
品牌: ADI [ ADI ]
 浏览型号3B37-B-06的Datasheet PDF文件第1页浏览型号3B37-B-06的Datasheet PDF文件第3页浏览型号3B37-B-06的Datasheet PDF文件第4页浏览型号3B37-B-06的Datasheet PDF文件第5页浏览型号3B37-B-06的Datasheet PDF文件第6页浏览型号3B37-B-06的Datasheet PDF文件第7页浏览型号3B37-B-06的Datasheet PDF文件第8页  
3B37  
GENERAL DESCRIPTION  
AC1310 ranging card, which houses user-supplied resistors to  
The 3B37 is a single-channel signal conditioning module that  
interfaces, amplifies and filters input voltages from a J, K, T, E, R,  
S or B-type thermocouple and provides isolated simultaneous  
voltage and current outputs linear with input voltage. High  
accuracy internal cold junction compensation and a predictable  
upscale open circuit indication provide a complete signal  
conditioning solution. To accurately measure low level signals  
in electrically noisy environments, +1500 V peak of galvanic  
transformer-based isolation with a common mode rejection  
(CMR) of 160 dB @ 50/60 Hz and a normal mode rejection  
(NMR) of 60 dB @ 50/60 Hz are provided. The current output  
withstands 130 V rms without damage and interfaces user  
equipment through screw terminals located on the 3B Series  
backplane. This plug-in, mix-and-match, hot-swappable module  
is easily field calibrated via front-panel zero and span  
determine zero and span. To facilitate selecting resistors, a  
Windows program, 3B-Custom, calculates resistor values based  
on the user-desired input/output ranges.  
Inside the 3B37 Module – A chopper-stabilized low-drift input  
amplifier assures stable long-term stability. At the amplifier  
input, a zero-scale input voltage is subtracted from the input  
signal to set the zero-scale value. For user convenience, the zero  
and span optionally can be factory configured to meet custom  
needs (Model 3B37-CUSTOM) or can be externally  
programmed (Model 3B37-X-00) via user-supplied resistors  
inserted in the optional AC1310 plug-on ranging card. Zero  
suppression can exceed 100% of the input range. This enables  
suppression of a zero-scale input value amny times larger that  
the total span for precise expanded-scale measurements of a  
selection portion of an input signal. The differential input  
circuit on the field side is fully floating, eliminating the need for  
any input grounding. Signal isolation by transformer coupling  
users a proprietary modulation technique for linear, stable and  
reliable performance. A demodulator on the computer side of  
the signal transformer recovers the original signa, which is then  
filtered and buffered to provide a low-noise, low-impedance  
output voltage, this output also drives a voltage-to-current (V/I)  
converter to provide a simultaneous current output for  
interfacing versatility.  
adjustments for both voltage and current outputs.  
Accurate and System Ready – Internal cold-junction  
compensation largely corrects errors arising from parasitic  
thermocouples formed by thermocouple connection to the  
input screw terminals, providing an accuracy of +0.5oC over the  
+5oC to +45oC ambient temperature range. The module  
generates a predictable upscale signal to indicate an open  
thermocouple; for a downscale response, connect a 47M, 0.25  
Watt resistor across screw terminals 2 and 4 on the 3B Series  
backplane  
3B Series Custom Ranging Program – Externally-  
programmable version Model 3B37-X-00, enable the user to  
configure a special input range by using the optional plug-on  
.
.
Figure 2  