10-15 Watts
xppower.com
CU10-15 Series
•
•
•
•
•
•
•
Low Cost
Small Size
PCB Mount
15 Watt Medical Version
Single Output 3.3-24 V
Peak Load Capability
Non-standard Outputs Available
Specification
Input
General
Input Voltage
Input Frequency
Input Current
• 85-264 VAC (120-370 VDC)
Efficiency
Isolation
• See tables
• 47-63 Hz
• 3000 VAC Input to Output (CU10/15)
4000 VAC Input to Output (CU15-M)
• 0.13 A rms at 230 VAC (CU10)
0.20 A rms at 230 VAC (CU15)
Switching Frequency
• 100 kHz typical for 10 W models
67 kHz typical for 15 W models
• 2.43 W/in3 (CU10); 3.17 W/in3 (CU15)
Inrush Current
• 20 A at 115 VAC, 40 A at 230 VAC,
cold start 25 ºC
Power Density
MTBF
Earth Leakage Current • Class 2 (no earth)
• >500 kHrs per MIL-HDBK-217F
(15 & 24 V units >400 kHrs)
Input Protection
• 1A fuse (CU10)
2A fuse (CU15, CU15-M)
Environmental
Output
Operating Temperature • CU10/15: 0 ˚C to +65 ˚C, derate from
full load at +45 ˚C to no load at +65 ˚C
Output Voltage
Output Voltage Trim
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
• See table
CU15-M: 0 ˚C to +70 ˚C, derate from
full load at +50 ˚C to 50ꢀ load at +70 ˚C
•
•
5ꢀ
1ꢀ
Cooling
• Convection-cooled
• 95ꢀ RH, non-condensing
• -20 ˚C to +85 ˚C
• 3000 m
• No minimum load required
• 1.5 s max
Operating Humidity
Storage Temperature
Operating Altitude
Vibration
• 14 ms max
• 16 ms typical at full load and 115 VAC
• 0.6ꢀ
• 10 Hz to 500 Hz, 2 g for 10 mins/cycle
60 min each cycle
Drift
Line Regulation
Load Regulation
Transient Response
• 0.5ꢀ max
• 1.0ꢀ max 10ꢀ load to full load
EMC & Safety
• 4ꢀ max deviation, recovery to within 1ꢀ
within 500 µs for 25ꢀ load change
Emissions
• CU15-M: EN55011 Level B conducted
Others: FCC20780 Level B,
Ripple & Noise
• 1ꢀ max pk-pk (see note 1)
EN55022 Class B conducted
Overvoltage Protection • 130-150ꢀ of Vnom, recycle input to reset
Short Circuit Protection • Trip and restart (Hiccup mode)
ESD Immunity
Radiated Immunity
EFT/Burst
• EN61000-4-2, level 3, Perf Criteria A
• EN61000-4-3, level 3, Perf Criteria A
• EN61000-4-4, level 2, Perf Criteria A
• EN61000-4-5, level 3, Perf Criteria A
• EN61000-4-6, 10 V, Perf Criteria A
Temperature
Coefficient
• 0.05ꢀ/˚C
Surge
Conducted Immunity
Dips & interruptions
• EN61000-4-11, 30ꢀ 10 ms,
60ꢀ 1000 ms, 100ꢀ 5000 ms,
Perf Criteria A, B, B
Safety Approvals
• CU15-M: EN60601, UL2601-1, CSA22.2
No. 601.1 per cUL
Others: EN60950, UL1950, CSA22.2 No.
234 per cUL
1