60 Watts  
AEM60 Series  
CEC 2008 & EISA 2007 Compliant 12 V  
Worldwide Medical Approvals  
4000 VAC Isolation  
Class II Construction  
Single Outputs from 5 V to 48 V  
High Efficiency  
3 Year Warranty  
Specification  
Input  
General  
Input Voltage  
• 80-264 VAC, derate output power 5%  
Efficiency  
• 85%, see note 5  
<90 VAC and 10% <85 VAC  
• 47-63 Hz  
• 1.5 A rms max  
• 80 A max at 240 VAC  
• Fitted with a T2 A/250 VAC fuse in live line  
Isolation  
• 4000 VAC Input to Output  
• 100 kHz typical  
• 4.2 W/Inch  
• 300 kHrs to MIL-HDBK-217F  
at 25 ˚C, GB  
Input Frequency  
Input Current  
Switching Frequency  
Power Density  
MTBF  
3
Inrush Current  
Input Protection  
No Load Input Power • <0.5 W for ≥12 V output  
Environmental  
Operating Temperature • 0 ꢀC to +60 ꢀC, derate linearly from 100%  
Output  
load at +40 ꢀC to 50% load at +60 ꢀC  
Cooling  
• Convection-cooled  
• 15-95% non-condensing  
Output Voltage  
Initial Set Accuracy  
Minimum Load  
• See table  
• See table  
• No mimimum load required  
• 3 s max  
• 3 ms  
Operating Humidity  
Storage Temperature • -20 ꢀC to +85 ꢀC  
Operating Altitude  
Vibration  
• 3000 m  
Start Up Delay  
• 5-500 Hz at 3 g for 10 mins on each axis  
• 30 g with 18 ms half sine wave,  
3 times on each axis  
Start Up Rise Time  
Hold Up Time  
Shock  
• 8 ms minimum at 115 VAC  
Line Regulation  
Load Regulation  
Transient Response  
1%  
• See table  
• 5% max deviation recovering to within 1%  
within 500 µs for 50% load change  
EMC & Safety  
Emissions  
• EN55011 Level B conducted & radiated  
• EN61000-3-2, class A  
• EN61000-3-3  
• EN61000-4-2 Level 3, Perf Criteria A  
• EN61000-4-3 Level 2, Perf Criteria A  
• EN61000-4-4, Level 3, Perf Criteria A  
• EN61000-4-5 Level 3, Perf Criteria A  
• EN61000-4-6 Level 3, Perf Criteria A  
Ripple & Noise  
• 1% max, 20 MHz bandwidth (see note 2)  
Harmonic Currents  
Voltage Flicker  
ESD Immunity  
Radiated Immunity  
EFT/Burst  
Overvoltage Protection • See table  
Overload Protection • 120-150%, trip & restart (hiccup mode),  
auto-recovery  
Short Circuit Protection • Continuous  
Temperature  
Coefficient  
0.05%/ꢀC  
Surge  
Conducted Immunity  
Dips & Interruptions  
• EN61000-4-11, 30% 10 ms, 60% 100 ms,  
100% 5000 ms Perf Criteria A, B, B  
Safety Approvals  
• UL60601-1, EN60601-1, IEC60601-1,  
CE Mark