欢迎访问ic37.com |
会员登录 免费注册
发布采购

ACM24US48 参数 Datasheet PDF下载

ACM24US48图片预览
型号: ACM24US48
PDF下载: 下载PDF文件 查看货源
内容描述: [AC-DC Power Supplies]
分类和应用:
文件页数/大小: 4 页 / 686 K
品牌: XPPOWER [ XP POWER LIMITED ]
 浏览型号ACM24US48的Datasheet PDF文件第1页浏览型号ACM24US48的Datasheet PDF文件第3页浏览型号ACM24US48的Datasheet PDF文件第4页  
AC-DC Power Supplies  
ACM24 Series  
General  
Characteristic  
Minimum  
Typical  
Maximum  
Units  
Notes & Conditions  
Typical average of efficiencies measured at 25%, 50%, 75%  
and 100% load and 115 VAC input  
Efficiency  
87  
%
Energy Efficiency  
Isolation  
Level VI  
4000  
VAC  
µA  
Input to Output, 2 x MOPP  
At 264 VAC, 60 Hz, from output to ground  
Variable  
Leakage Current  
Switching Frequency  
Mean Time Between Failure  
Weight  
30  
100  
70  
24  
kHz  
kHrs  
lb (g)  
250  
MIL-HDBK-217F at 25 °C GB  
Body only  
0.35 (160)  
Environmental  
Characteristic  
Operating Temperature  
Storage Temperature  
Operating Humidity  
Operating Altitude  
Cooling  
Minimum  
Typical  
Maximum  
+60  
Units  
Notes & Conditions  
0
-20  
5
°C  
°C  
%
m
Derate from 100% load at 40 °C to 50% load at 60 °C  
+85  
90  
RH, non-condensing  
5000  
Natural convection  
Shock  
1 m drop onto concrete on each of 6 axes, non operating  
2 g, 0.3 decades/min, 15 mins for each of 3 axes  
Vibration  
10  
300  
Hz  
EMC: Emissions  
Phenomenon  
Conducted  
Standard  
Test Level  
Notes & Conditions  
EN55032  
EN55032  
Level B  
Level B  
Class A  
Radiated  
Harmonic Current  
Voltage Flicker  
EN61000-3-2  
EN61000-3-3  
EMC: Immunity  
Phenomenon  
Standard  
Test Level  
Criteria  
as below  
Notes & Conditions  
Medical Device EMC  
Low Voltage PSU EMC  
ESD Immunity  
IEC60601-1-2  
EN61204-3  
Ed.4.0 : 2014  
High severity level  
8 kV contact, 15 kV air  
10 V/m  
as below  
EN61000-4-2  
EN61000-4-3  
EN61000-4-4  
EN61000-4-5  
EN61000-4-6  
EN61000-4-8  
A
A
A
A
A
A
B
B
B
A
B
B
B
Radiated Immunity  
EFT/Burst  
Level 3  
Surge  
Level 3  
Conducted Immunity  
Magnetic Fields  
6 V  
30 A/m  
Dip: 100% 10 ms  
Dip: 70% 500 ms  
Int: 100% 5000 ms  
Dip: 30% 25 AC Cycles  
Int: 100% 0.5 AC Cycle  
Int: 100% 1 AC Cycle  
Int.: >95% 5000 ms  
EN61000-4-11  
Dips and Interruptions  
At 8 angles  
EN60601-1-2  
Safety Approvals  
Safety Agency  
Safety Standard  
UL60950-1  
Notes & Conditions  
UL  
TUV  
CB  
EN60950-1  
Information Technology  
Medical, 2 x MOPP  
IEC60950-1  
CE  
LVD  
UL  
ANSI/AAMI ES 60601-1  
CSA C22.2 No. 60601  
EN60601-1  
CSA  
TUV  
CB  
IEC60601-1  
www.xppower.com  
2