RF3619D
• 402 to 405 MHz Medical Band Front-end Filter
• Single-ended Input/Outout Operation
• Complies with Directive 2002/95/EC (RoHS)
Pb
404.0 MHz
SAW Filter
Value
+10
3
-10 to +60
-40 to +85
260
Absolute Maximum Ratings
Rating
Input Power Level
DC Voltage on any Non-ground Pin
Operating Temperature Range
Storage Temperature Range in Tape and Reel
Solder Reflow Temperature, 10 seconds, 5 cycles maximum
Units
dBm
V
°C
°C
°C
SM3838-8 Case
3.8 x 3.8
Electrical Characteristics
Characteristic
Center Frequency at 25 °C
Insertion Loss
Passband Ripple, 402-405 MHz
3 dB Bandwidth
Attenuation relative to IL
MIN
:
398.5 MHz
408.5 MHz
383.5 MHz
423.5 MHz
Freq. Temp. Coefficient
Frequency Aging, First Year
Single-ended Input/Output Impedance Match
Differential Input/Output Impedance Match
Lid Symbolization (Y=year WW=week S=shift)
Standard Reel Quantity
Reel Size 7 Inch
Reel Size 13 Inch
9
FTC
|fA|
5
1
1
1, 3
30
25
40
38
35
45
50
48
-37
10
50
150
A15, YWWS
500 Pieces/Reel
3000 Pieces/Reel
ppm/k
ppm
ohms
ohms
dB
BW
3
Sym
f
c
IL
MIN
Notes
1, 2, 3
1, 3
1, 3
1
Minimum
Typical
404.0
2.1
0.6
Maximum
2.3
0.8
Units
MHz
dB
dB
MHz
5.0
6.0
CAUTION: Electrostatic Sensitive Device. Observe precautions for handling.
Notes:
1.
2.
3.
4.
5.
6.
7.
8.
9.
Unless noted otherwise, all measurements are made with the filter installed in the specified test fixture which is connected to a 50
test system.
The frequency f
c
is defined as the midpoint between the 3 dB frequencies.
Where noted specifications apply over the entire specified operating temperature range of -10 °C to +60 °C.
Frequency aging is the change in fc with time and is specified at +65 °C or less. Aging may exceed the specification for prolonged temperatures
above +65 °C. Typically, aging is greatest the first year after manufacture, decreasing significantly in subsequent years.
The design, manufacturing process, and specifications of this device are subject to change.
One or more of the following U.S. Patents apply: 4,54,488, 4,616,197, and others pending.
All equipment designs utilizing this product must be approved by the appropriate government agency prior to manufacture or sale.
Tape and Reel Standard Per ANSI / EIA 481.
This product complies with Directive 2002/95/EC of the European Parlament and of the Council of 27 January 2003 on the restriction of the use of
certain hazadous substances in electrical and electronic equipment.
www.RFM.com E-mail: info@rfm.com
© 2011-2012 by RF Monolithics, Inc.
Page 1 of 3
RF3619D - 5/10/12