欢迎访问ic37.com |
会员登录 免费注册
发布采购

CW1251M 参数 Datasheet PDF下载

CW1251M图片预览
型号: CW1251M
PDF下载: 下载PDF文件 查看货源
内容描述: 800-2500VA可编程交流电源系列 [800-2500VA PROGRAMMABLE AC POWER SERIES]
分类和应用:
文件页数/大小: 5 页 / 256 K
品牌: POWERBOX [ Powerbox ]
 浏览型号CW1251M的Datasheet PDF文件第1页浏览型号CW1251M的Datasheet PDF文件第2页浏览型号CW1251M的Datasheet PDF文件第3页浏览型号CW1251M的Datasheet PDF文件第5页  
CW Series  
800-2500VA PROGRAMMABLE AC POWER SERIES  
CW SPECIFICATION  
MANUAL (M)  
OUTPUT  
PROGRAMMABLE (P)  
POWER  
CW 801  
800VA  
1250VA  
CW 1251  
CW 2501  
2500VA  
Power factor of load  
Phase  
0 lag to 0 lead  
All models single phase output  
VOLTAGE  
Ranges  
0-135VRMS or 0 to 275VRMS, user selectable (consult factory for custom output voltages)  
+/-0.1% of range < 100Hz, +/-0.2% of range,> 100Hz (>5  
VAC out)  
Accuracy  
+/-1% of range (> 5VAC out)  
Resolution  
0.1VRMS  
0.25% typical < 100Hz add 0.5%/100Hz above 100Hz  
Total harmonic distortion  
AC noise level  
< 50 mVRMS typical for CW801 & CW1251, 100 mVRMS typical for CW2501  
Amplitude stability  
+/-0.1% of full scale over 8 hours at constant line, load and +/-0.05% of full scale over 8 hours at constant line, load and  
temperature after 15 minute warm-up typical  
temperature after 15 minutes warm-up typical  
Load regulation  
Line regulation  
+/-0.1% of full scale voltage for a full resistive load to no load  
(<10 mVRMS typical, measured at point of sense)  
+/-0.1% of full scale voltage for a +/-10% line change from nominal line voltage  
(<5 mVRMS typical, measured at point of sense)  
Remote voltage sense  
Compensation for up to 5 VRMS total lead voltage drop  
CURRENT  
CW 801  
6.0 ARMS in 135VAC range or 3.0 ARMS in 270 VAC range  
9.4 ARMS in 135VAC range or 4.7 ARMS in 270VAC range  
CW 1251  
CW 2501  
Accuracy  
Resolution  
18.6 ARMS in 135VAC range or 9.3 ARMS in 270VAC range  
+/-0.5% typical  
0.1 ARMS  
+/-0.6% max  
0.01 ARMS  
FREQUENCY  
Range  
45 to 500Hz(45 to 1000Hz optional on programmable unit only)  
Accuracy  
+/-0.5% typical  
0.1Hz  
+/-0.02% max  
Resolution  
0.1Hz, 0.01Hz for remote programming  
PHASE ANGLE  
Range  
-359 to +359 degrees. Positive phase indicates time lag from  
reference  
Accuracy  
Within 100 microseconds of equivalent angle  
1 degree  
Resolution  
MEASUREMENT  
Voltage range  
Accuracy  
0-270 VRMS  
+/-1% of full range for voltage (>5VAC out), measured at  
point of sense  
+/-0.1% of range < 100Hz, +/-0.2% of range >100Hz,  
measured at point of sense (> 5VAC out)  
Resolution  
0.1 VRMS  
Current*  
Range  
CW 801  
0 to 6.0 ARMS  
0 to 9.4ARMS  
0 to 18.6ARMS  
CW 125I  
CW 2501  
Accuracy  
+/-2% of range for linear loads with current  
0.2A for 800VA & 1200VA, 0.4A for 2500VA  
+/-0.6% of range for linear loads with current  
0.2A for 800VA & 1200VA, 0.4A for 2500VA  
Resolution  
0.1ARMS  
0.01 ARMS  
Note: In a parallel system, the current/power displayed in the master unit is the sum of all units in the system.  
Your dependable power partner – www.powerbox.com.au  