2
Wide Range Test Voltage Settings
Features
■
Generate Test Voltages Across a Wide Spectrum
The
can generate test voltages ranging from 250 V
3455
to 5 kV. Settings can be made in steps as fine as 25 V. Very
high insulation resistance measurement up to 5 teraohms is
possible.
Ideal for All Insulation Diagnostic Applications
Functions such as automatic calculation and display of PI
(Polarization Index) and DAR (Dielectric Absorption Ratio),
as well as step voltage test, temperature compensation,
temperature measurement, and leakage current display make
the 3455 suitable for a variety of diagnostic applications.
Data Memory Function
The 3455 provides a manual storage function for 100 data
and a logging function for 10 data (360 times). The date and
time of measurement are also recorded.
safety regulations for category IV measurements (600
V). A shutter mechanism prevents simultaneous use of
measurement terminals and other terminals. Other safety
features include a voltage measurement
USB Interface
Easily transfer data to a PC
via the USB interface using
our free PC application
function, high-voltage warning
indicator, and auto discharge function.
software. The software also
features a convenient report
Large, Easy to Read Display
The display is backlit and features a
creation function.
logarithmic bar graph as an analog
Safety Foremost
The 3455 complies with
type indicator in addition to the digital
readout
.
Primary Measurement Functions
■
Insulation resistance measurement
Leakage current display
❚
❚
Measurement voltage is selectable from 250 V, 500 V, 1.00
kV, 2.50 kV, and 5.00 kV. More finely graded settings are also
possible. When measurement is completed, the unit shows
the insulation resistance value, test voltage (setting and actual
output), leakage current, DAR, PI, and elapsed time.
When measuring insulation resistance, the instrument can be
switched to display leakage current. This is possible before,
during, and after measurement.
Display of leakage current during measurement
Actual output
voltage
Insulation resistance measurement display
I
nsulation
Actual output
voltage
resistance value
> flashes
if less than 1 nA
> flashes if
measurement
range was
exceeded
Elapsed time
leakage current value
STEP 5.0 kV step voltage test
Elapsed time
Insulation resistance value
ꢋꢁꢌꢍꢎꢏꢁꢐꢁꢑꢀ
ꢒꢓꢐꢂꢔꢁꢀꢁꢕ
ꢈꢉꢊ
�ꢀꢁꢂꢈ
Step voltage test
In this type of test, the voltage is gradually raised and the insulation resistance
❚
ꢇꢉꢊ
ꢆꢉꢊ
ꢅꢉꢊ
ꢃꢉꢊ
�ꢀꢁꢂꢇꢄ
�ꢀꢁꢂꢆꢄ
and leakage current change is measured. Two different step settings are
�ꢀꢁꢂꢅꢄ
available: 500 V 1 kV
1.5 kV
2 kV
2.5 kV and 1 kV
2 kV
3
→
→
→
→
→
→
�ꢀꢁꢂꢃꢄ
kV 4 kV 5 kV. The test time for each step can also be selected.
→
→